• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET LEAD SMALL-S, 86 CM; NO MATCH Back to Search Results
Model Number 1456Q/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/18/2024
Event Type  Injury  
Event Description
It was reported that the patient presented to the hospital for a scheduled procedure.Postoperative check showed that the left ventricular (lv) lead was dislodged.The lv lead was explanted and replaced on (b)(6) 2024.The patient was in stable condition.
 
Manufacturer Narrative
The reported events were lead dislodgement and failure to capture.As received, a complete lead was returned in one piece.The reported event of failure to capture was not confirmed.Visual examination of the lead did not find any anomalies.Electrical testing of the lead did not find any indication of conductor fractures or internal shorts.S-curve was measured to be within specification.Dhr reviewed and found to be complete.The product passed all quality control tests prior to its distribution.Correction: section d6b: date of explant is updated, and it is (b)(6) 2024.
 
Event Description
New information received notes that the left ventricular (lv) lead exhibited loss of capture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET LEAD SMALL-S, 86 CM
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18872692
MDR Text Key337290052
Report Number2017865-2024-34525
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510158
UDI-Public05414734510158
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1456Q/86
Device Lot NumberA000148957
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/18/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received03/13/2024
Supplement Dates FDA Received04/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-