DEPUY IRELAND - 9616671 ATTUNE CONV SHIM SZ5 10MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Catalog Number 254500928
Device Problems Crack (1135); Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 03/05/2024
Event Type  malfunction  
Event Description
It was reported that the shims were cracked, missing metal connecting pieces.
 
Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: it was reported, that the shims were cracked/missing metal connecting pieces.Patellar clamp is sticking/not functioning properly.The device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the returned sample revealed, that there was no damage or defects with the attune conv shim.Based on the drawing specifications, a new design solution was developed for the device attune conv shim and the solution involves.Now it is going to be a single piece/all plastic part.Thus get rid of the metal bushing.The overall complaint was unconfirmed.As the observed, condition of the attune conv shim would not contribute to the complained device issue.Based on the investigation findings, it has been determined, that no corrective and/or preventative action is proposed.There is no indication, that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV SHIM SZ5 10MM
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18875077
MDR Text Key337534129
Report Number1818910-2024-05538
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295135050
UDI-Public10603295135050
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254500928
Device Lot NumberBFA1KNS
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received04/09/2024
04/11/2024
Supplement Dates FDA Received04/10/2024
04/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-