MEDTRONIC HEART VALVES DIVISION EVOLUT PRO PLUS VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVPROPLUS-26 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/07/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: product id d-evprop23-29 (lot: 0011966159); product type: 0195-heart valves; product id l-evprop23-29 (lot: 0011924339); product type: 0195-heart valves; medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the valve dislodged due to angulated aorta and dilated ascending aorta.Subsequently a second valve was implanted.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received that during the valve implant, a pre-implant balloon aortic valvuloplasty (bav) was performed.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Updated data: h6 conclusion: images were submitted to medtronic for review.The excerpt of the image subject matter expert (sme) review report follows: intra procedural fluoroscopic images were provided for review of the event description.Patient¿s executive summary was not provided for anatomical review.The valve was deployed just prior to the point of no recapture; however, an angiogram was not performed thus proper depth assessment could not be validated.The valve appeared to have dislodged for unknown reasons.It is not possible to determine cause of dislodgement.Subsequently, the dislodged valve was snared, and a second valve was successfully implanted.The intra procedural images shows a possible dissection int the right external iliac artery; however, it was not reported and there is no evidence that any intervention was performed.As stated in the ifu, potential risks associated with the implantation of the evolut pro+ bioprosthesis may include, but are not limited to, the following: prosthetic valve migration/embolization; vascular access-related complications (for example, dissection, perforation, pain, bleeding, hematoma, pseudoaneurysm, irreversible nerve injury, compartment syndrome, arteriovenous fistula, stenosis).Potential factors that can influence a dislodged valve include tension applied on the delivery catheter system (dcs) during positioning, calcification levels in the native vessel, compliance of the aorta and native vessels, and a number of others, but a root cause could not be established from the information available.This event does not indicate device misuse or malfunction.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|