Model Number EVPROPLUS-26
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2024
Event Type  Injury  
Manufacturer Narrative
Continuation of d10: product id d-evprop23-29 (lot: 0011966159); product type: 0195-heart valves; product id l-evprop23-29 (lot: 0011924339); product type: 0195-heart valves; medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during implant of this transcatheter bioprosthetic valve, the valve dislodged due to angulated aorta and dilated ascending aorta.Subsequently a second valve was implanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Additional information was received that during the valve implant, a pre-implant balloon aortic valvuloplasty (bav) was performed.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Manufacturer Narrative
Updated data: h6 conclusion: images were submitted to medtronic for review.The excerpt of the image subject matter expert (sme) review report follows: intra procedural fluoroscopic images were provided for review of the event description.Patient¿s executive summary was not provided for anatomical review.The valve was deployed just prior to the point of no recapture; however, an angiogram was not performed thus proper depth assessment could not be validated.The valve appeared to have dislodged for unknown reasons.It is not possible to determine cause of dislodgement.Subsequently, the dislodged valve was snared, and a second valve was successfully implanted.The intra procedural images shows a possible dissection int the right external iliac artery; however, it was not reported and there is no evidence that any intervention was performed.As stated in the ifu, potential risks associated with the implantation of the evolut pro+ bioprosthesis may include, but are not limited to, the following: prosthetic valve migration/embolization; vascular access-related complications (for example, dissection, perforation, pain, bleeding, hematoma, pseudoaneurysm, irreversible nerve injury, compartment syndrome, arteriovenous fistula, stenosis).Potential factors that can influence a dislodged valve include tension applied on the delivery catheter system (dcs) during positioning, calcification levels in the native vessel, compliance of the aorta and native vessels, and a number of others, but a root cause could not be established from the information available.This event does not indicate device misuse or malfunction.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO PLUS VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18875332
MDR Text Key337329925
Report Number2025587-2024-01515
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeAR
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEVPROPLUS-26
Device Catalogue NumberEVPROPLUS-26
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received03/27/2024
04/12/2024
05/15/2024
Supplement Dates FDA Received03/28/2024
04/15/2024
05/20/2024
Date Device Manufactured11/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient SexMale
-
-