MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Stopped (1503)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Manufacturer Narrative
The investigation is ongoing.A getinge technician will investigate the hl 20.A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the hl20 pump intermittently displayed the error message: ¿runaway¿.The event occurred during a routine check.No harm to any person has been reported.Complaint id: (b)(4).
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
It was reported that the hl20 pump intermittently displayed the error message: ¿runaway¿.The event occurred during a routine check.No harm to any person has been reported.A getinge field service technician (fst) was sent for investigation and repair on 2024-03-29.The tacho strobe foil was replaced.And as a preventive measure all connection of the pump were reset and the belt tension re-adjusted.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.Additional device observed were made to make sure the failure is solved on 2024-03-18 and 2024-03-29.As confirmed by the getinge field service technician during repair on 2024-03-29 the most probably root causes was determined as defect marks on the strobe foil.Some of the marks were no longer clearly visible.Since the sensor no longer detects one or more marks on the tacho strobe foil and it incorrectly measures a low speed at the pump head.The review of the non-conformities has been performed on 2024-04-18 for the period of 2007-10-21 to 2024-02-28.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2007-10-21.Based on the results the reported failure "error message: ¿runaway¿" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18876327
MDR Text Key337357259
Report Number8010762-2024-00132
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701043262
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received04/16/2024
Supplement Dates FDA Received04/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2007
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-