Catalog Number 72200616
Device Problems Corroded (1131); Overheating of Device (1437); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/19/2024
Event Type  malfunction  
Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Event Description
It was reported that during a rotator cuff repair/ acromioplasty, the mdu started making alarming sounds and became extremely hot to the touch.The procedure was completed with a smith and nephew back up device.There was a delay less than 30 minutes and no further complications were reported.
 
Manufacturer Narrative
Internal complaint reference (b)(4).The reported device was received for evaluation.A visual inspection found no physical damage on the exterior of the product.A functional evaluation was performed on the returned device and found a noisy motor.Further evaluation revealed a corroded gearbox.It was also noted that the device motor house began to get hot.These failures have been determined to be related to the reported event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with a mechanical component failure.Factors that could have contributed to the failure include gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18876419
MDR Text Key337552797
Report Number1643264-2024-00167
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received04/02/2024
Supplement Dates FDA Received04/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-