• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL COBALT¿ HF QUAD CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL COBALT¿ HF QUAD CRT-D MRI SURESCAN¿; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTPB2QQ
Device Problems Failure to Deliver Shock/Stimulation (1133); Device Sensing Problem (2917)
Patient Problem Tachycardia (2095)
Event Date 01/27/2024
Event Type  Injury  
Manufacturer Narrative
Continuation of d10: 407645 lead, implanted (b)(6) 2023. 638bl32 heart band, implanted (b)(6)2009.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) inappropriately withheld therapy for ventricular tachycardia (vt) that was detected as supra ventricular tachycardia (svt).The crt-d remains in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
It was further reported that a new wavelet was obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT¿ HF QUAD CRT-D MRI SURESCAN¿
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ   1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18877199
MDR Text Key337353363
Report Number9614453-2024-00986
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000178178
UDI-Public00763000178178
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDTPB2QQ
Device Catalogue NumberDTPB2QQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2024
Initial Date FDA Received03/11/2024
Supplement Dates Manufacturer Received03/14/2024
Supplement Dates FDA Received04/02/2024
Date Device Manufactured09/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
439878 LEAD, 6935M62 LEAD
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-