Model Number MI1200 SYNCHRONY |
Device Problem
Impedance Problem (2950)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 10/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
Hearing performance with the device is affected.Affected channels were noticed during in situ measurements.Further information has been requested.
|
|
Manufacturer Narrative
|
Additional information: based on the received information from the field, a damage to the active electrode likely caused by device minute mobility is suspected.However, to determine an exact root cause device investigation at the explanted device is necessary.No information on possible further steps has been received.
|
|
Event Description
|
The user_s hearing performance with the device is affected.No decision regarding re-implantation has been made so far.
|
|
Search Alerts/Recalls
|