MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  Injury  
Manufacturer Narrative
H3, "other" code: the device was discarded at the facility.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
It was reported to gore a gore® cardioform septal occluder was selected to treat a patent foramen ovale (pfo).Having deployed the left disc in the left atrium without issues, the physician reportedly started to deploy the right disc in the right atrium without realizing that a portion of the right disc was also deployed in the left atrium.Due to the now unequal deployment forming a somewhat elongated shape, it was not possible to lock the device and it prolapsed into the left atrium.It was therefore decided to remove the device.To do so, the physician opened the retrieval cord lock but, reportedly in error, pulled the handle until the entire cord was removed from the handle.As the device now no longer was connected to the retrieval cord, it embolized into the aorta.Subsequently, the device was caught with a snare and removed from the patient.To proceed with the procedure, an amplatzer® occluder was selected and the pfo was closed successfully.Reportedly, the patient was doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
thomas giebing
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18882876
MDR Text Key337403201
Report Number2017233-2024-04701
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-