• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number C40+
Device Problem Impedance Problem (2950)
Patient Problem Failure of Implant (1924)
Event Date 03/07/2024
Event Type  malfunction  
Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
Until (b)(6) 2023 the user had obtained always good benefit from the device.Since then, the user noticed a decline in the hearing performance of the device.The user requested the explantation of the implant.No new device was implanted.
 
Manufacturer Narrative
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
 
Event Description
Until june 2023 the user had obtained always good benefit from the device.Since then, the user noticed a decline in the hearing performance of the device.The device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18883800
MDR Text Key337765151
Report Number9710014-2024-00222
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberC40+
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received05/06/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age30 YR
Patient SexFemale
-
-