Model Number C40+ |
Device Problem
Impedance Problem (2950)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 03/07/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
Until (b)(6) 2023 the user had obtained always good benefit from the device.Since then, the user noticed a decline in the hearing performance of the device.The user requested the explantation of the implant.No new device was implanted.
|
|
Manufacturer Narrative
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
|
Event Description
|
Until june 2023 the user had obtained always good benefit from the device.Since then, the user noticed a decline in the hearing performance of the device.The device was explanted.
|
|