Model Number MMT-105ELPKNA |
Device Problem
Computer Software Problem (1112)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/28/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
Information received by medtronic indicated that the customer reported that the inpen was not logging the correct doses of insulin.Troubleshooting was performed and found that the intended dose amount and dose amount recorded in the inpen app was greater than 1.0 unit.It was found that the customer reported dose knob/dial was difficult to turn.No harm requiring medical intervention was reported.The customer will discontinue the use of the inpen and will be returned for analysis.
|
|
Manufacturer Narrative
|
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Cartridge holder locks properly in place.Missing dose window was noted during visual inspection.The inpen paired to the commercial app.Inpen passed baseline and wireless functionality.App logbook displayed: 14.5u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u, 15u.Inpen passed displacement dose accuracy.Inpen passed dialing alignment using dose knob zero alignment gage.Inpen passed front cap investigation.Inpen was received with missing dose window.In conclusion: inpen passed all required testing.No anomalies were noted and all functions tested ok.Dose log/report data inaccuracy was not confirmed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|
|