It was reported before use, during a routine check performed by the customer, the cardiosave intra-aortic balloon pump (iabp) had the following errors recorded 4 times: power up fault code # 13 and fault code # 101.System failed to boot up and produced a loud audible alarm.There was no patient involvement.
|
Updated fields: b4, d9(device available for eval), g3, g6, h2, h3, h4, h6(investigation type, investigation findings & investigation conclusions), h10 a getinge field service engineer (fse) was dispatched to evaluate the unit.Fse was able to confirm the reported issue in the unit logs.The fse disassembled the unit.Entire system was vacuumed and cleaned.All pcbas were removed, cleaned and reseated correctly.System rebooted.Verified in configuration data that all boards were recognized and reporting properly.The unit passed all functional and safety tests.
|