It was reported, the probe tip of the subject device fell off inside the patient's body during a therapeutic vascular resection of the liver parenchyma resulting to no output.The fallen part was retrieved, and the procedure was completed with a similar device.At the time of the incident, the device was set to seal & cut 3, cut 3 mode with intelligent monitoring turned on.A probe check was conducted prior to use.There were no reports of patient harm.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was evaluated by olympus, and it was confirmed the probe was broken at 17 mm from the distal end of the probe.Furthermore, there were non-reportable (non-pae) defects noted.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the probe broke which led to the recovery of the subject device occurred due to one of the following: 1) during the output activation in seal & cut mode, the distal end of the probe was slightly contacting a thin equipment, causing spark generation.2) a force was applied to the probe during spark generation.3) cracks occurred in the probe due to the load applied to the probe during tissue grasping and seal & cut output.4) due to continuous use of the device, the cracks on the probe progressed.This led to breakage of the probe.However, a specific root cause of the reported event could not be identified.The event can be prevented by following the instructions for use which state: "do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities." "if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure." olympus will continue to monitor field performance for this device.
|