Model Number A4674A
Device Problems Material Twisted/Bent (2981); Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and the evaluation found that the defect was shown to be caused by the customer¿s mishandling.There are no additional reportable malfunctions.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the rigid endoscope telescope was slightly bent and had a defective optic.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.The event likely occurred due to excessive force by the customer.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 3 MM, 30°, WIDEANGLE, AUTOCLAVABLE
Type of Device
RIGID ENDOSCOPE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg/de/germany
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18886218
MDR Text Key337802637
Report Number9610773-2024-00684
Device Sequence Number1
Product Code GCP
UDI-Device Identifier04042761006361
UDI-Public04042761006361
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K904939
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA4674A
Device Lot Number800284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/06/2024
Initial Date FDA Received03/12/2024
Supplement Dates Manufacturer Received03/19/2024
Supplement Dates FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-