SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE BLOOD AND FLUID WARMER; WARMER, THERMAL, INFUSION FLUID
|
Back to Search Results |
|
Model Number HL-90 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing, there was a continuous low water alarm.There was no patient involvement and no patient harm adverse event reported.
|
|
Manufacturer Narrative
|
B3: date of event is unknown; no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
H6.Investigation codes: updated.Investigation summary: the affected device was returned for evaluation in good condition.After visual inspection, put water in water tank, attached disposable temp check, plugged line cord into outlet and connected to electrical analyzer and turned device on.The alarm that sounded was for disposable and not for low water.The customer's complaint was not replicated.However, the root cause for the disposable alarm was due to the microswitch missing the lever arm.Replaced microswitch.Device passed all functional tests after repair.The service history review identified there was no indication that the complaint was related to a service of the device within the review period.
|
|