This report is being supplemented to provide additional information based on the legal manufacturer's final investigation results.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 1 year since the subject device was manufactured.Based on the results of the investigation, the broken probe was likely caused by contacting other surgical instruments, or the non-insulated areas as described in the mechanism below.1.The wearing of the tissue pad likely occurred because ¿seal & cut¿ output was activated while grasping nothing with the grasping section and the probe tip.Or the user continued to activate the output after a transection of tissue.2.The non-insulated area of the grasping section and the probe came into contact due to wearing of the tissue pad.3.The output in ¿seal & cut¿ was activated while the non-insulated area of the grasping section was in contact with the probe.As a result, a contact mark developed.4.A force to activate the output in ¿seal &cut¿ mode or a force to grasp tissue was applied to the probe.Therefore, cracks developed at the contact mark.5.A force was applied to the probe causing it to break.However, the root cause of the broken broke could not be determined.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ ¿when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation.¿ this supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.Also, information has been added to h4.Olympus will continue to monitor field performance for this device.
|