Model Number ICO-3000
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Manufacturer Narrative
The customer has not indicated that she had a pcr confirmation of her covid diagnosis.No lot number provided,unable to effectively verify and confirm customer feedback.A false negative result may occur if the test result is read before 15 minutes or after 30 minutes.There was no indication to confirm or deny if the user/patient had utilized the test kit appropriately as per intended use or off use.
 
Event Description
Event details((b)(4)): i ordered/received the 5-pack of covid tests, because i had tested positive at a pharmacy and needed to be able to ascertain my health status to rejoin society.To test the product you sent me, i used it right after my (+) test, as a control, assuming a (+) test result.No, it tested (-).So, i used another of your tests, and one more.All 3 gave false negative results while i was actively infected, sick in bed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IHEALTH
Type of Device
Coronavirus antigen detection test system.
Manufacturer (Section D)
ANDON MEDICAL CO.,LTD
tianjin, airport economic area 30030 8
CH  300308
Manufacturer (Section G)
ANDON MEDICAL CO.,LTD
tianjin, airport economic area 30030 8
CH   300308
Manufacturer Contact
ren jun
tianjin, airport economic area 30030-8
CH   300308
MDR Report Key18891608
MDR Text Key337757331
Report Number3008573045-2024-00007
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EUA210470
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial
Report Date 03/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberICO-3000
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/27/2024
Initial Date FDA Received03/12/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age35 YR
-
-