Returned product consisted of an express sd balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 7mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found a pinhole in the balloon.
|
Reportable based on the device analysis completed on (b)(6) 2024.It was reported that balloon leak occurred.The patient was diagnosed with fibrosing mediastinitis and pulmonary vein stenosis.Interventional treatment with pulmonary vein stent placement was performed under general anesthesia.The lesion site was mildly tortuous and moderately calcified.After placing an 8f non-boston scientific (bsc) guide catheter, a 0.14 guidewire was used to cross the lesion.Following balloon pre-dilatation, a 7.0mmx19mmx150cm express vascular sd stent was advanced for treatment.However, during inflation at 4 atmospheres, the image showed that the contrast agent leaked from the stent balloon tip.After withdrawal, visual inspection found liquid leakage, but no visible pinhole was noted.The procedure was completed with another of same device.There were no complications reported and the patient status was stable.However, returned device analysis revealed a pinhole 7mm from the tip.
|