Unomedical reference number (b)(4) event occurred in canada on (b)(6) 2024, it was reported that the infusion set's tubing got detached at the tubing connector while the patient was sleeping.The blood glucose level of the patient was 15.6 mmol/l.The site location was patient's abdomen, and the pump was located on the same side.The infusion had been used for three to four hours.Reportedly, the infusions were not stored, or used, in a place where they might have been exposed to extreme temperatures and humidity.There was no stress or pull on the tubing and the pump was not dropped with the set connected to patient's body.No further information was available.
|