PERFUSION SYSTEMS CARDIOBLATE; SURGICAL DEVICE, FOR CUTTING, COAGULATION, AND/OR ABLATION OF TISSUE, INCLUDING
|
Back to Search Results |
|
Model Number 49260 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluation summary: visual inspection shows one of the jaw tips is broken off.Reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a cardioblate gemini, the customer reported that after completing the first set of ablations which happened to be on the right hand side concave jaws, the customer noted once the 10 ablations were completed that the tip of the jaw had broken off.The end of the jaw was still attached to the guide that was in the patient.The customer was able to remove the the broken tip by pulling the guide out of the chest and removing the broken tip.The device was replaced to complete the procedure and used with the original guides without further incident.There was no adverse patient effect associated with this event.Medtronic received additional information that the jaws of the device clamped onto a surface 10 times before the device broke besides tissue.The jaws were closed and locked 10 times.The broken tip did not fall into the patient.It was attached to the guide while in the patient.
|
|
Search Alerts/Recalls
|
|
|