The lens and used cartridge were returned loose in the opened lens pouch.An opened company cartridge 10-count carton was also returned with nine unopened company cartridge.One sample was pulled randomly for evaluation.The lens had viscoelastic dried on it.The lens was cleaned with klrp for further evaluation.One haptic was cracked distal area.The optic had two long cracks from the edge to the center.The cracks were angled opposite of each other.This damage may indicate a plunger override occurred.Other smaller cracks and scratches were observed, which appear to have been caused by an instrument used to grasp the lens (parallel marks on posterior and anterior surface).The used company cartridge was evaluated.Viscoelastic was observed in the cartridge.The cartridge tip has heavy stress.The bottom of the cartridge tip was torn/ripped with a damage portion extending past the tip.This extensive damage would indicate the lens/plunger were not in acceptable positions.Excessive force would have been necessary to cause this type of damage.The used company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.One of the returned unopened samples was pulled for evaluation.The company cartridge was microscopically examined with no damage or abnormalities observed.The cartridge was functionally tested per the dfu (directions for use) using a qualified company iii blue handpiece, sn60wf, 27.0 diopter lens and viscoelastic.The cartridge was filled with viscoelastic per the dfu.The lens was loaded and biased down.The lens was advanced making sure the plunger was in the correct position in contact with the trailing optic edge.No lens or cartridge damage was observed after the lens delivery.The company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The root cause for the observed lens damage could not be determined.The optic had two long cracks from the edge to the center.These may have been interpreted as the reported scratch.The cracks were angled opposite of each other.This damage may indicate a plunger over/underride occurred.Other smaller cracks and scratches were observed, which appear to have been caused by an instrument used to grasp the lens (parallel marks on posterior and anterior surface).The returned used company cartridge was severely damaged.The bottom of the cartridge tip was torn/ripped with a damage portion extending past the tip.This extensive damage would indicate the lens/plunger were not in acceptable positions.Excessive force would have been necessary to cause this type of damage.The used company cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.One of the returned unopened company cartridges was evaluated.Functional and dye stain testing was conducted with the unopened sample with acceptable results.No lens or cartridge damage was observed.Top coat dye stain testing was conducted with acceptable results.The type of damage observed to the lens and cartridge may occur when the lens and plunger are not in proper positions for advancement.The ifu(instructions for use) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd (ophthalmic viscosurgical device)-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.Handpiece ifu: important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact alcon.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The manufacturer internal reference number is: (b)(4).
|
A health care professional reported that during an intraocular lens (iol) implant procedure, scratched lens, noted plastic fragments inside the cartridge after lens was inserted.Patient was left aphakic.Additional information was requested, but no further information is available.
|