Model Number 1458QL/86
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2024
Event Type  malfunction  
Event Description
During an initial implant procedure on (b)(6) 2024, it was reported that the physician had difficulties removing the stylet from the left ventricular (lv) lead, and when more force was applied, the insulation of the lead became damaged as a result.The lv was not installed, and a new lv lead was successfully implanted.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported events of stylet could not be removed and lead damaged were confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.Visual inspection of the lead found that the connector pin and connector cap were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage and the ptfe stylet coating was bunched up/clogged with the inner coil distal to the connector pin.The cause of the reported event was due to bunched up ptfe coating of the stylet that prevented the removal of the stylet and excessive forces resulted in the connector pin and cap to be pulled out of the connector assembly.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18894840
MDR Text Key337556335
Report Number2017865-2024-34871
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458QL/86
Device Lot NumberA000148033
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/28/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received04/23/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age81 YR
Patient SexMale
Patient Weight95 KG
-
-