Catalog Number H938737 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/18/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an exactamix 250 ml eva (ethylene vinyl acetate) tpn (parenteral nutrition) bag had a black particle inside the filling port connector.This was observed after removing the empty bag from the overwrap, prior to connecting to the compounder.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H4: the lot was manufactured between august 4, 2023 and august 8, 2023.H10: the device was received for evaluation.During visual inspection a dark particulate matter was observed on the inner wall of the fill port connector.Under magnification the particulate was noted to be dried dark unknown particle matter that was noted to be attached to the inner wall edge area of the fill port connector in fluid pathway.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|