MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535FCS
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.The probe was broken at 10 mm from the distal end of the probe.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The event can be detected/prevented by following the instructions for use (ifu) which state: ·"do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.¿ when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." based on the results of the investigation the exact cause of the event could not be determined.The likely mechanism causing the reported event could have been the following: 1.The tissue pad was worn out because the seal and cut was output with the gripper closed (including after the tissue was cut) without grasping the tissue.2.Since the tissue pad was worn out, non-insulated area of the grasping section and the distal end of the probe came into contact.3.The output was activated in seal & cut mode in state of description stated above.Therefore, scratches (contact marks) were made on the probe and the grasping section, indicating that they came into contact with each other.4.The device was activated in seal &cut mode or while the grasping section was grasping tissue.This applied a force to the scratched area of the grasping section, causing this area to have cracks.5.A force was applied to the probe causing it to break.Olympus will continue to monitor field performance for this device.
 
Event Description
It was reported during a total laparoscopic hysterectomy, the active blade of the subject device broke and fell off.Items that fell off have been collected.The procedure was completed with a similar device.No health hazards were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Type of Device
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18895699
MDR Text Key337741294
Report Number9614641-2024-00648
Device Sequence Number1
Product Code GEI
UDI-Device Identifier04953170409677
UDI-Public04953170409677
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K211838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTB-0535FCS
Device Lot Number31K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-