Model Number G172
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/05/2024
Event Type  Injury  
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Device replacement was recommended, and a replacement interval was discussed.No adverse patient effects were reported.This device remains in service.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Device replacement was recommended, and a replacement interval was discussed.No adverse patient effects were reported.This device remains in service.If information is provided in the future, a supplemental report will be issued.Additional information was received that surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The return of the device has been requested.This report will be updated upon return and completion of analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOGEN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18897360
MDR Text Key337644563
Report Number2124215-2024-15286
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/13/2017
Device Model NumberG172
Device Catalogue NumberG172
Device Lot Number481087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/05/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received03/22/2024
Supplement Dates FDA Received03/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age72 YR
-
-