ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA MP ICD; NO MATCH Back to Search Results
Model Number CD3369-40Q
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/29/2024
Event Type  Injury  
Event Description
It was reported that the patient presented with an implantable cardioverter defibrillator (icd) that was exhibiting mechanical breakdown.The icd was explanted, and new icd was implanted.The patient condition was unknown.
 
Manufacturer Narrative
Further information was requested but not received.
 
Manufacturer Narrative
The reported event of mechanical breakdown could not be confirmed.Interrogation of the device revealed the device was at elective replacement indicator (eri) when received.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.A longevity calculation was performed and found the battery depletion was normal based on the device usage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18898854
MDR Text Key337599797
Report Number2017865-2024-34881
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Model NumberCD3369-40Q
Device Lot NumberA000038454
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/22/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received04/19/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA; QUARTET; TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient Age81 YR
Patient SexFemale
-
-