• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SAFESTEP HUBER NEEDLE SET 20G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SAFESTEP HUBER NEEDLE SET 20G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problem Activation Problem (4042)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Event Description
It was reported that during needle removal, the base did not push down in the middle of the needle and could not activate the safety mechanism.There was no reported patient injury.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of difficulty advancing the safety mechanism was confirmed and the cause was determined to be supplier related.The product returned for evaluation was one 20g safestep infusion set.The investigation findings were consistent with misplaced or excess manufacturing adhesive, which bound the safety mechanism to the needle.The following observations were made during the sample evaluation: ¿ the safety mechanism was not engaged over the needle tip and was returned positioned at the needle base.Microscopic examination under uv light assistance revealed a white/clear adhesive-like substance at the interface of the needle shaft and safety mechanism.That material fluoresced under ultraviolet light, which was consistent with the adhesive used to assemble the infusion set.From this, it appeared that adhesive was deposited on the needle shaft during device manufacture.The device is a supplied component and the supplier was notified of the event.H3 other text : evaluation findings are in section h11.
 
Event Description
It was reported that during needle removal, the base did not push down in the middle of the needle and could not activate the safety mechanism.There was no reported patient injury.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 20G X 0.75IN
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18899136
MDR Text Key337664532
Report Number3006260740-2024-01064
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741066207
UDI-Public(01)00801741066207
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0031
Device Lot NumberASHSFC014
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/13/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-