Model Number 364510
Device Problems Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/18/2023
Event Type  malfunction  
Event Description
An orsiro drug-eluting stent system was selected for treatment of the lad.During the coronary angioplasty procedure, there was dislodgement between the balloon and the stent during an attempt to overcome a lesion in the middle third of the lad.
 
Manufacturer Narrative
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested by means of manufacturing process output monitoring.Based on the conducted investigations, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 4.0/26
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18899275
MDR Text Key337633249
Report Number1028232-2024-01394
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier07640130411753
UDI-Public(01)07640130411753(17)2501
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P170030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number364510
Device Catalogue NumberSEE MODEL NO.
Device Lot Number01230311
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/13/2024
Initial Date FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/23/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-