Model Number 5620006 |
Device Problem
Material Puncture/Hole (1504)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/13/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The holes in bag would disrupt the patient therapy, if not caught prior to use.The clinician would need to replace the circuit during the procedure causing an interruption in patient therapy.
|
|
Event Description
|
Holes in the breathing bags.
|
|
Manufacturer Narrative
|
The holes in bag would disrupt the patient therapy, if not caught prior to use.The clinician would need to replace the circuit during the procedure causing an interruption in patient therapy.The defect of hole in the breathing bags was confirmed.Holes were identified along the parting line near the neck of the breathing bag.Dcr-14039 has been initiated as a result of the findings.Airlife will continue to monitor and trend complaints.
|
|
Event Description
|
Holes in the breathing bags.
|
|