Model Number C40+ |
Device Problem
Circuit Failure (1089)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 02/27/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user's hearing performance with the device is affected.Intermittent hearing is reported.
|
|
Event Description
|
The user's hearing performance with the device is affected.Intermittent hearing is reported.
|
|
Manufacturer Narrative
|
Additional information: according to the complaint information and the manufacturer_s experience with this kind of device, it is assumed that the device might have failed due to humidity ingress at the housing braze joint through micro-leaks.However, to determine an exact root cause of failure a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
|
Search Alerts/Recalls
|