MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number SONATA
Device Problem Impedance Problem (2950)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Event Description
In situ testing showed affected channels, however, the recipient did not report an important decrease in hearing benefit with the device.The speech audiometry did show a decrease.The recipient reported not having benefit with the implant.However, no follow-up measurements are available.
 
Manufacturer Narrative
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event Description
In situ testing showed affected channels, however, the recipient did not report a big decrease in hearing benefit with the device.The speech audiometry did show a decrease.The recipient reported not having benefit with the implant.However, no follow-up measurements are available.
 
Manufacturer Narrative
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However, to determine an exact root cause a device investigation of the explanted device is necessary.Reportedly the recipient no longer has hearing benefit with the device.However no information on possible further steps has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key18901185
MDR Text Key337747456
Report Number9710014-2024-00240
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/01/2011
Device Model NumberSONATA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/07/2024
Date Device Manufactured06/01/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age52 YR
Patient SexFemale
-
-