A customer alleged discrepant results for two patient samples tested with the elecsys ft4 ii assay on the cobas 8000 e 602 module.Sample 1 resulted in a ft4 value of 2.15 ng/dl when tested on the e 602 module.The result was higher than expected and was repeated on the same e 602 module, resulting in a ft4 value of 1.17 ng/dl.Sample 2 resulted in a ft4 value of 2.45 ng/dl when tested on the e 602 module.The result was higher than expected and was repeated on the same e 602 module, resulting in a ft4 value of 1.52 ng/dl.The repeat results were deemed correct.
|
The serial number of the cobas e 602 module is (b)(6).Due to repeated calibration failures, the customer replaced the reagent pack, calibrated, and ran qc which was acceptable.The field service engineer (fes) confirmed there was an issue with the specific reagent pack the customer had replaced.The fes performed checks and verified the instrument by having the customer perform calibration, qc, and performing patient correlations successfully.The investigation determined the replacement of the reagent pack resolved the issue.
|