Model Number SPL-T |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/12/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that the subject device cable connector has been broken.The issue occurred during preparation for use for percutaneous nephrolithotomy procedure.The procedure was completed with a similar device.There were no reports of patient harm.
|
|
Manufacturer Narrative
|
The device was returned and investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
|
Event Description
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The device was returned to olympus for evaluation and the customer's allegation was confirmed.Device evaluation found damaged connector.In addition, device evaluation found scratches and dings on the housing.A definitive root cause of the phenomenon cannot be conclusively identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.
|
|