• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. SHOCKPULSE LITHOTRIPSY TRANSDUCER.; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-T
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Event Description
It was reported that the subject device cable connector has been broken.The issue occurred during preparation for use for percutaneous nephrolithotomy procedure.The procedure was completed with a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned and investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event Description
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The device was returned to olympus for evaluation and the customer's allegation was confirmed.Device evaluation found damaged connector.In addition, device evaluation found scratches and dings on the housing.A definitive root cause of the phenomenon cannot be conclusively identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18903854
MDR Text Key337665442
Report Number3011050570-2024-10036
Device Sequence Number1
Product Code FEO
UDI-Device Identifier00821925043831
UDI-Public00821925043831
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2024
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-