MEDTRONIC HEART VALVES DIVISION EVOLUT FX VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number EVOLUTFX-29 |
Device Problems
Device Dislodged or Dislocated (2923); Patient Device Interaction Problem (4001); Migration (4003)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/06/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic has requested additional information pertaining to this reportable event.If additional reportable information is received, a supplemental report will be submitted. medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that 7 days following implant of this transcatheter bioprosthetic valve, the valve migrated.The valve was interfering with the mitral valve so it was decided to attempt to snare the valve and move it out of the mitral valve.Upon the attempt to snare the valve, the valve subsequently dislodged.A new medtronic valve was implanted without complication.It was noted that the implant depth of the valve contributed to the migration.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received that the valve dislodged ventricular and then aortic during the attempted snare.Prior to valve dislodgement, the implant depth on the non-coronary cusp (ncc) was 6 millimeter¿s (mm) and the implant depth on the left coronary cusp (lcc) was 8 mm.After the valve dislodged, the implant depth on the ncc was 3 mm and the implant depth on the lcc was 4 mm.The patient had a large left ventricular outflow tract (lvot) that contributed to the dislodge.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Corrected data: a3b.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Manufacturer Narrative
|
Corrected: h6 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|