Catalog Number 08324131190 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The serial number of the e602 analyzer is (b)(6).Calibration signals recovered within expectations.Controls recovered acceptably.The investigation is ongoing.
|
|
Event Description
|
The initial reporter stated they received a questionable result for one patient sample tested with elecsys folate iii on a cobas e 602 module in comparison to the result measured using a previous version/lot of the folate assay.The sample initially resulted in a folate value of 6.83 ug/l when tested using folate reagent lot 726792.The customer repeated the patient sample as part of comparison studies for the new folate reagent, lot 726840.The repeat result with the new lot was 5.06 ug/l.
|
|
Manufacturer Narrative
|
The customer's calibration and control data did not suggest a local or general reagent issue with either of the assay versions/lots.The customer's qc recovery is comparable to their peers.Internal method comparison data, as well as data from standardization, do not suggest a systemic bias between the assay versions.The investigation did not identify a product issue.The cause of the event could not be determined.
|
|
Search Alerts/Recalls
|