COVIDIEN MFG DC BOULDER LIGASURE XP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number LXMJ37L |
Device Problems
Device Alarm System (1012); Energy Output Problem (1431)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Event Description
|
According to the reporter, during procedure, the device was connected to a generator with software version 4.0.4 and showed an error code of incomplete seal was being given prior to activation of energy.Another device was used using the same generator and it worked successfully.There was no patient injury.
|
|
Manufacturer Narrative
|
D10 concomitant product: vlft10gen - vlft10gen ft series energy platformx1, serial# unknown medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: b5, d9 (return date), g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted no abnormalities.Functionally, the device's jaw opening and closing mechanism was functioning properly.The weld integrity of the device was inspected and was found to be within specification.The knife cut of the device was tested on a silicone test strip with acceptable results.The heel gap of the device was measured and it was found to be within specification.The product analysis found the device produced an instant regrasp alarm when activation attempts were made.The device was disassembled and it was identified that the red wire had disconnected from the connector on the rotation wheel.It was reported that there was an issue with the alarm activation.The reported issue was confirmed.The most likely cause was traced to device design.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
According to the reporter, during total hysterectomy, the device was connected to a generator with software version 4.0.4 and showed an error code of incomplete seal was being given prior to activation of energy while trying to seal the connective tissues when beginning to mobilize but couldn't engage.There were no completed seals.There was no bleeding.Cleaning was very limited since it wasn't able to be used.Another device was used using the same generator and it worked successfully.There was no patient injury.
|
|
Event Description
|
According to the reporter, during total hysterectomy, activation and end tone were heard but sealing was inadequate/partial.There was an error code of incomplete seal prior to activation of energy while trying to seal the connective tissues when beginning to mobilize but could not engage.There were no completed seals.There was no bleeding.Cleaning was very limited since it was not able to be used.Another device was used using the same generator and it worked successfully.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|