• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458QL/86
Device Problems Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that the patient presented to the hospital for a scheduled procedure.The patients left ventricular (lv) lead was noted to have no capture threshold due to dislodgement.The lv lead was explanted and replaced on (b)(6) 2024.The patient was in stable condition throughout.
 
Manufacturer Narrative
The reported events were lead dislodgement and failure to capture.As received, a complete lead was returned in one piece.The reported event of failure to capture was not confirmed.Visual inspection, electrical testing and x-ray examination of the lead found no conductor fractures or internal shorts and no anomalies.The s-curve hump height was measured to be within product specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18910021
MDR Text Key337728701
Report Number2017865-2024-35062
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734510219
UDI-Public05414734510219
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458QL/86
Device Lot NumberA000119619
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received05/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GALLANT ICD; RIGHT ATRIAL CONDTENDRIL; RIGHT VENTRICULAR OPTISURE LEAD
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-