SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Event Description
It was reported that b30 and e216 scope communication errors were observed using the light source.The issue occurred during an unspecified procedure.There were no reports of patient harm.
 
Manufacturer Narrative
The device was not returned.Troubleshooting was performed advising to power off both the cv-190 and clv-190, remove the scope, clean the contact points on the scope with an alcohol preparation pad, wait for the scope to dry, reconnect the scope, power everything back up and observe the result.In addition, it was advised to always disconnect or connect the scope when the light source is powered off if b30 or e216 errors are present.Both the cv-190 and clv-190 must be powered off.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.New information added to the following field: d9, h3, and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over nine (9) years since the subject device was manufactured.The device was not returned to olympus; instead, the customer requested a visit from a field service engineer (fse) to the facility.The fse reported that the device was inspected, and no issues were verified.Based on the results of the investigation, a definitive root cause could not be determined.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18911639
MDR Text Key337744033
Report Number3002808148-2024-02452
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CV-190 - EVIS EXERA III VIDEO SYSTEM CENTER.
-
-