Model Number FSESL |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.Patient demographics not obtained at this moment.
|
|
Event Description
|
As reported, with this foresight sensor measurement showed 90 on the forehead when the value expected was 70-80.No error message was displayed.It was tried different area of the same patient, and even on different persons.There was no allegation of patient injury.The device was available for evaluation.Patient demographics not obtained.
|
|
Manufacturer Narrative
|
One foresight sensor was received for a full evaluation.The report of measurement issue was not able to be confirmed.No visible damage was observed from the sensor during visual examination.Sensor monitored tissue oxygen saturation (sto2) on hemosphere monitor without any error message.The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.No root cause could be identified for the issue since no defect was found during the product evaluation.
|
|