• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES FORESIGHT LARGE SENSOR; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES FORESIGHT LARGE SENSOR; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number FSESL
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2024
Event Type  malfunction  
Manufacturer Narrative
Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.Patient demographics not obtained at this moment.
 
Event Description
As reported, with this foresight sensor measurement showed 90 on the forehead when the value expected was 70-80.No error message was displayed.It was tried different area of the same patient, and even on different persons.There was no allegation of patient injury.The device was available for evaluation.Patient demographics not obtained.
 
Manufacturer Narrative
One foresight sensor was received for a full evaluation.The report of measurement issue was not able to be confirmed.No visible damage was observed from the sensor during visual examination.Sensor monitored tissue oxygen saturation (sto2) on hemosphere monitor without any error message.The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.No root cause could be identified for the issue since no defect was found during the product evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORESIGHT LARGE SENSOR
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
MDR Report Key18912843
MDR Text Key337764970
Report Number2015691-2024-02040
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K223865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Model NumberFSESL
Device Lot Number41202248
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2024
Initial Date FDA Received03/15/2024
Supplement Dates Manufacturer Received05/06/2024
Supplement Dates FDA Received05/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-