Catalog Number 04404483190 |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Low Test Results (2458)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/21/2024 |
Event Type
malfunction
|
Event Description
|
We received an allegation about discrepant results for 1 patient's serum sample tested with sodium (na) test and 1 patient's serum sample tested with na test and glucose hk gen.3 (gluc3) assay on a cobas 6000 c501 analyzer.Sample 1: na: initial result: 165 mmol/l.Repeat result: 140 mmol/l.Sample 2: na: initial result: 150 mmol/l.Repeat result: 141 mmol/l.Gluc3: initial result: 42 mg/dl.Repeat result: 106 mg/dl.The medical personnel questioned the initial results and requested the samples be repeated.The repeat results were deemed to be correct.
|
|
Manufacturer Narrative
|
The na electrode lot number and expiration date were not provided.The gluc3 reagent lot number was 74348401.The expiration date was not provided.The field service engineer found that the rinse vacuum line was ripped at the nozzle.He repaired the vacuum line and inspected the rinse tubing and the rinse levels.The customer performed qc and mechanical checks and they were successful.Precision studies were performed and they were within specifications.The investigation is ongoing.
|
|
Manufacturer Narrative
|
Section d4 was updated.Correction: sample 1 and sample 2 were plasma samples.The glucose calibration was last performed on (b)(6) 2023 and it was acceptable.The sodium calibration was performed on (b)(6) 2024 and it was acceptable.Qc recovery was within the range.No indication of a performance issue with the reagent.The service action (repairing the vacuum line and inspecting the rinse tubing and the rinse levels) resolved the issue.No further issues were reported afterward.
|
|
Search Alerts/Recalls
|