• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07P70-20
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2024
Event Type  malfunction  
Manufacturer Narrative
Complete information for section a patient information, 1.Patient identifier = sid= (b)(6).All available patient information was included.Additional patient details are not available.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed falsely elevated alinity i free t4 (ft4) results which does not match with clinical picture for multiple patients.The result provided were: on 26feb2024 patient 1: sid (b)(6) initial = 22.94 pmol/l /repeated=13.15 pmol/l additional information provided: ft3=3.65 pmol/l; tsh=3.147 uiu/ml patient 2: initial=19.54 pmol/l /repeated on 27feb2024=11.90 pmol/l and 11.73 pmol/l on 14mar2024 53yrs old male patient initial=20.37 pmol/l /repeated=14.82 pmol/l additional information provided: tsh=2.816 miu/ml laboratory reference range for ft4= 9.0 to 19.0 pmol/l there was no reported impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I FREE T4 REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18913759
MDR Text Key337777987
Report Number3005094123-2024-00122
Device Sequence Number1
Product Code CEC
UDI-Device Identifier00380740121815
UDI-Public00380740121815
Combination Product (y/n)N
Reporter Country CodeBK
PMA/PMN Number
K173122
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07P70-20
Device Lot Number57272UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-