• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND DIAGNOSTICS DIVISION ALINITY I FREE T4 REAGENT KIT; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 7P70-77
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2024
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient information has been included.No additional patient details are available.
 
Event Description
The customer observed falsely elevated alinity i free t4 results for one patient.The initial test result of ft4 was 64.35 pmol/l, and the retest result of the same sample was 12.8 pmol/l.No impact to patient management was reported.
 
Manufacturer Narrative
Corrected data found in section d4 catalogue # and lot #.Data and information provided by the customer were reviewed and support the complaint issue.An increase in complaints has been observed for lot 56201ud02, however, in-house performance testing was completed which indicates the product is performing as expected.Device history review did not identify any non-conformances or deviations with the list number and complaint issue.In-house performance testing was completed which indicates the product is performing as expected.Labeling was reviewed which adequately addresses the current issue.Based on the investigation, no systemic issue or deficiency of the alinity i free t4 assay was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I FREE T4 REAGENT KIT
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
ABBOTT IRELAND DIAGNOSTICS DIVISION
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
nicole jenne
max-planck-ring 2
post market surveillance
wiesbaden 65205
GM   65205
6122582960
MDR Report Key18915130
MDR Text Key337797554
Report Number3005094123-2024-00124
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K173122
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7P70-77
Device Lot Number56201UD02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY I PROCESSING MODU, 03R65-01, AI02805.; ALNTY I PROCESSING MODU, 03R65-01, AI02805.
-
-