Catalog Number D134801 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/19/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and three hours into the catheter usage while creating roof line after pvi (pulmonary vein isolation), when the stsf was removed from the patient's body and flushed, the irrigation hole was found to be clogged.No error messages occurred.The catheter was replaced and the procedure continued.The procedure was completed without patient's consequence.
|
|
Manufacturer Narrative
|
E1 initial reporter phone: (b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
On 20-mar-2024, the product investigation was completed as the complaint device was not returned.Since no device has been received for analysis, no product investigation can be performed, and the customer complaint cannot be confirmed.A manufacturing record evaluation was performed for the finished device number lot 31180003l and no internal action related to the complaint was found during the review.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|