Updated fields - b4,d9,e1(event site telephone : (b)(6).G3,g6,h2,h3,h6(type of investigation,investigation findings,component code,investigation conclusions),h10.Getinge field service engineer (fse) found on march 5, 2024, went to the customer's site and found that the loop leakage was caused by the failure of the safety disk and the drive valve group.We gave the hospital a quotation and the customer agreed to repair it.On april 24, 2024, fse went to the site to replace the drive valve group and the safety disk, and performed the factory-specified tests.All passed, met the clinical use requirements, and were delivered to the customer for use.Patient involved and no personal injury occurred.
|