• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2024
Event Type  malfunction  
Manufacturer Narrative
H11: section a: through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak is confirmed and was determined to be use related.Two 22g x 3/4" powerloc max infusion sets were returned for evaluation.The returned product samples were evaluated and a tear in the extension tube was seen at the interface of the proximal luer adapter.The fracture surfaces of the damage contained striation-like patterns which were indicative of flexural fatigue based failures, which may have been a contributing factor to the extension tubing damage.Repetitive mechanical stresses such as twisting and kinking may have contributed to the observed event; however, it appeared that additional unidentified factors also contributed.The device is a supplied component and the supplier has been notified of this event.This complaint will be recorded for future trending and monitoring purposes.
 
Event Description
It was reported by customer that, "we've started seeing a trend at facility with needle.We've had repeated instances cracked tubing with leaking." no other information was provided.A specific number of affected devices or patients was not provided by the complainant.Two devices were returned for evaluation.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18916367
MDR Text Key337809541
Report Number3006260740-2024-01138
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741047497
UDI-Public(01)00801741047497
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0142275
Device Lot NumberASGXFC025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/26/2024
Initial Date FDA Received03/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-