Catalog Number 530.705 |
Device Problems
Physical Resistance/Sticking (4012); Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/05/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device would not run and had a sticky trigger.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.(b)(4).
|
|
Event Description
|
It was reported from switzerland that during service and evaluation, it was determined that the battery/reamer drill device would not run, the mode switch resistance was low, the device had a sticky trigger, and it was difficult to assemble/disassemble.It was further determined that the device failed pretest for check cannulation, check for sticky trigger, check the function of the device, and check the power with the power test bench.It was noted in the service order that the device did not function.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|