ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/22/2024
Event Type  malfunction  
Event Description
During the shift check, the autopulse platform (b)(6) displayed a fault code 08 (motor controller fault detected) error message.No patient involvement.
 
Manufacturer Narrative
The reported complaint that the autopulse platform (b)(6) displayed a fault code 08 (motor controller fault detected) was confirmed during functional testing and review of the archive data.The root cause was due to a failed drivetrain motor, likely attributed to the age of the device.The autopulse platform was manufactured in (b)(6) 2014 and is 10 years old, well past its expected serviceable life of 5 years.During visual inspection, there was no physical damage observed on the autopulse platform.During the archive data review, a fault code 08 was observed, thus confirming the customer complaint.The autopulse platform failed initial functional testing due to fault code 08, thus confirming the customer complaint.The drivetrain motor was replaced to remedy the fault.Following the service, the autopulse platform passed the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there was no similar complaint reported for the autopulse platform with (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim sackrison
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key18919228
MDR Text Key337844331
Report Number3010617000-2024-00231
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2024
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-