Model Number SONATA |
Device Problem
Impedance Problem (2950)
|
Patient Problem
Failure of Implant (1924)
|
Event Date 03/08/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
In early (b)(6) 2023, the parents reported a decline in hearing performance with the left implant.The recipient had no obvious auditory response when wearing the left external device.The parent denied a history of trauma.No swelling or redness was seen.The recipient has been re-implanted.
|
|
Manufacturer Narrative
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
|
|
Event Description
|
In early december 2023, the parents reported a decline in hearing performance with the left implant.The recipient had no obvious auditory response when wearing the left external device.The parent denied a history of trauma.The recipient has been re-implanted.
|
|