GYRUS ACMI, INC. GYRUS, PK-SP GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 744000
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2024
Event Type  malfunction  
Event Description
It was reported that the generator did not fulgurate but would only resect.It was displaying checks for irrigation and electrodes.The event was found during preparation for use.There were no reports of patient harm.
 
Manufacturer Narrative
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation and device evaluation results.New information added to the following fields: b5, d9, h3, h4, h6 and h10.Correction added in fields: d3, g1 and facility number should be 3011050570 instead of 3003790304 in the mfr.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over twenty (20) years since the subject device was manufactured.The device was returned to olympus for inspection, and the reportable malfunction was not confirmed.Based on the results of the investigation, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Event Description
The customer reported the intended procedure was diagnostic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYRUS, PK-SP GENERATOR
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
GYRUS ACMI, INC.
800 west park drive
brooklyn park MN 01581
Manufacturer (Section G)
GYRUS ACMI, INC.
800 west park drive
westborough MA 01581
Manufacturer Contact
todd brill
800 west park drive
brooklyn park, MA 01581
5082077661
MDR Report Key18920771
MDR Text Key337859729
Report Number3003790304-2024-00021
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009271
UDI-Public00821925009271
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number744000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/29/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer Received04/11/2024
Supplement Dates FDA Received04/18/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-