U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG INLEXA 3 HF-T DF4 IS-1 PROMRI; CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG INLEXA 3 HF-T DF4 IS-1 PROMRI; CRT-D Back to Search Results
Model Number 404700
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that the device was explanted due to eri status.Should additional information be received, this file will be updated.Please note this icd is affected by a field safety corrective action, bio-lqc, initiated in march 2021.
 
Manufacturer Narrative
Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.There was no indication of a material or manufacturing problem.Subsequently the device was interrogated.The interrogation could be properly performed and revealed the eri battery status.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.There was no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INLEXA 3 HF-T DF4 IS-1 PROMRI
Type of Device
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key18921062
MDR Text Key337864106
Report Number1028232-2024-01425
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P050023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model Number404700
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2024
Initial Date FDA Received03/18/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-