Model Number MMT-105NNGYNA
Device Problems Break (1069); Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/12/2024
Event Type  malfunction  
Manufacturer Narrative
H6: type of investigation code-10 investigation findings code-758 investigation conclusions -140 = bc07af per visual inspection: there is physical damage to cartridge holder, broken pocket clip and inpen front shell does not attach in place.Missing injection button was noted.The inpen paired to the commercial app.Unable to perform displacement dose accuracy, baseline and wireless functionality due to broken off injection button.Inpen passed dialing alignment using dose knob zero alignment gage.Inpen passed front cap investigation.However, inpen was received with broken cartridge holder.In conclusion: unable to confirm dose log inaccuracy due to broken off injection button making it difficult to dispense.In addition, broken pocket clip and broken cartridge holder were noted during visual inspection.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event Description
Information received by medtronic indicated that the customer reported cosmetic damage and the clip on the cap broke.Troubleshooting was performed and found that the inpen app was not recording the exact doses dialed on the in pen.No harm requiring medical intervention was reported.The customer will discontinue using the inpen and the inpen was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INPEN MMT-105NNGYNA NOVO NORDISK GRAY
Type of Device
SYRINGE, PISTON
Manufacturer (Section D)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer (Section G)
COMPANION MEDICAL INC
1223 world trade drive, 100
san diego CA 92128
Manufacturer Contact
lusine boyadzhyan
1223 world trade drive, 100
san diego, CA 92128
MDR Report Key18922899
MDR Text Key337889225
Report Number3012822846-2024-00355
Device Sequence Number1
Product Code FMF
UDI-Device Identifier0000108620880003
UDI-Public(01)0000108620880003(17)231004(10)B0714
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201337
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/04/2023
Device Model NumberMMT-105NNGYNA
Device Catalogue NumberMMT-105NNGYNA
Device Lot NumberB0714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Date Manufacturer Received03/07/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age59 YR
Patient Weight81 KG
-
-